Service Manual PhysioFlow Lab1
16 / 20
Version of Monday, 10 July 2017
5.
Maintenance, Transport, Storage and Disposal
Service and maintenance
There is no operator or customer serviceable part, and no part has to be changed under normal tear and
wear conditions.
If the system does not work properly, please refer to the troubleshooting section. If the problem remains,
please contact your authorised dealer, or the customer support +33 3 72 82 50 00 or
.
With special authorisation of the manufacturer or the dealer, the unit’s embedded processor software can
be upgraded. To perform this operation, it is necessary to replace the EPROM.
First stage is to open the enclosure. To do that, please slide the 4 caches on the upper side of the device,
and unscrew.
Replacing the EPROM requires pliers, protected against static electricity. The EPROM must be retrieved
very slowly and carefully from the support, by gripping the edges of the EPROM. Be careful, the new
EPROM has to be put in place very carefully.
IT IS INDISPENSABLE THAT THE NEW EPROM IS POSITIONED THE SAME WAY THAN THE PREVIOUS ONE: THE
NOTCH ON THE UPPER SURFACE MUST BE DIRECTED TOWARDS THE POWER SUPPLY (SAME DIRECTION AS
THE PHYSIOFLOW® LOGO ON THE FRONT PANEL AND OPPOSITE TO THE PATIENT CONNECTOR).