M
anatec
B
iomedical
21, Rue du stade, Petit Ebersviller
▪
57730 Folschviller
▪
France
Tél : +33 (0)3 72 82 50 00
▪
Fax : +33 (0)1 30 74 46 48
Service Manual PhysioFlow Lab1
19 / 20
Version of Monday, 10 July 2017
Appendix B: Specifications
General
Dimension
343 x 260 x 84 mm
Weight
4.0 Kg
Protection
Full EMI protection, (Metallic enclosure with special EMC shielding)
Power supply
100-240 VAC 50/60 Hz
Fuses
1A/250V only
Power input
50 VA
Galvanic
insulation
of
patient, and optical coupling
Device of BF type
Electrodes
6 Ag/AgCl pregelled adhesive electrodes per examination
Recommended : Agilent HP40493E
Connectors
Back panel
Female SUB-D9 (serial link to PC)
Switcher-filter-fuses module Schaffner
Front panel
Specific patient connector (6 leads + ground)
Gilded contacts
Measurement current
Intensity of 3.6 mA peak to peak
Sinus frequency : 75 KHz
Operating temperature
5°C - 55°C
Patient cable
6 leads, EMCD shielded
Length 3.5 meters (ref. SP090)
or 10 meters (ref SP090-10M)
Gilded contacts
Specific RS232 serial link
Null modem type wiring (1,8 Meters)
SUB-D9M/SUBD9F connectors
Optical cable available, (for connection with a desktop computer).
Power supply cable
2P+T/CEE 22 (computer type) (3 meters)
Computer environment
Minimal
computer
configuration
PC Pentium compatible Desktop or Laptop
200 MHz, 32Mb RAM, HD 50 Mb free, screen 800x600, 65536 colours
Microsoft Windows 95 / 98 / Me / NT / 2000 operating system.
1 available serial link (RS 232)