Analog output pin connection and instructions
SUB-D 15 female connector on back panel.
Pin15 : Z
Pin 8 : Gnd
Pin13 : Stroke Volume
Pin 6 : Gnd
Pin11 : ECG
Pin 4 : Gnd
Pin9 : CO
Pin 2 : Gnd
The 4 outputs are -2.5V to +0.0V full scale / 10Kohms output impedance.
Please
connect
PhysioFlow®
to
a
medical
system
compliant
with
CE
regulations
or with your local medical regulations.
The
Stroke
Volume
and
Cardiac
Output
trends
only
operate
when
calibration
phase
is
completed
on
the
computer.
The
external
data
acquisition
system
has
to
be
calibrated
(adjustment
of
scales)
to
reproduce
measurements
displayed
on the PhysioFlow® Computer screen.
To perform this task, start a measurement on a volunteer who is very relaxed and stable. Please calibrate
SV and CO measurement on your acquisition system in accordance with the readings you obtain on the
PhysioFlow® screen. Then make a change (for instance by retrieving the white electrode), and calibrate
again your system in accordance with the new PhysioFlow® readings obtained.
When using PhysioFlow® Software 2.0.0 and later, only the Z (Pin15) and
ECG (Pin11) outputs are active. In this case, it not possible yet to get
neither CO on Pin9 nor SV on Pin13.
This feature is planned for a future software version. If you need it, please
contact the technical support:
.