32
LUCAS
®
3 Chest Compression System
Instructor Guidebook
2. Adjust the height of the Suction Cup to set the Start Position. This is the position
where the LUCAS device will start its two-inch compressions, and the point where it
will return the chest for full recoil.
Make sure the LUCAS device is in
ADJUST
mode
Push the Suction Cup down with two fingers until the Pressure Pad touches the
patient’s chest without compressing it
Push
PAUSE
to lock the start position.
Check for proper position. If you need to reposition:
Push
ADJUST
, pull the Suction Cup up, move the device by pulling the
Support Legs, push the Suction Cup down until the Pressure Pad touches
the chest, then push
PAUSE
once back in place
Push either
ACTIVE (continuous)
or
ACTIVE (30:2)
.
Notes:
§
The LUCAS device has a “quick fit” feature, so if the Pressure Pad is pushed down
too hard, or not touching the chest fully, the device will adjust the Pressure Pad by up
to 30 mm /1.2 inches in either direction to the correct start position.
§
Some users will use a marker to draw a line on the chest around the top and/or
bottom of the Suction Cup to help monitor placement during operation.
The LUCAS Stabilization Strap helps secure the correct position during operation. Apply it
while the LUCAS device is active to keep interruptions to a minimum. Delay the application
if it delays any medical treatment of the patient. The Stabilization Strap is not a neck
Stabilization Strap but a device Stabilization Strap. The Stabilization Strap can fit and be put
on the outside of a c-collar.
Moving the Patient
When you move the patient, you can secure the patient’s arms by placing their wrists in the
Patient Straps on the LUCAS device. This makes it easier to move the patient, but DO NOT
lift the patient by the Patient Straps. Make sure that IV access is not obstructed.
Summary of Contents for LUCAS 3
Page 2: ...2 LUCAS 3 Chest Compression System Instructor Guidebook ...
Page 4: ......
Page 6: ......
Page 8: ......
Page 9: ...Chapter 1 Instructor Training Preparation ...
Page 10: ......
Page 14: ......
Page 15: ...Chapter 2 Classroom Training Introduction and Overview ...
Page 16: ......
Page 36: ......
Page 37: ...Chapter 3 Small Group Hands on Training and Application Practice ...
Page 38: ......
Page 47: ...3 SMALL GROUP HANDS ON TRAINING AND APPLICATION PRACTICE 47 ...
Page 48: ......
Page 49: ...Chapter 4 Training Evaluation ...
Page 50: ......
Page 52: ......
Page 53: ...Appendix A Important Contacts ...
Page 54: ......
Page 56: ......
Page 57: ...Appendix B Frequently Asked Questions ...
Page 58: ......
Page 66: ......
Page 67: ...Appendix C The LUCAS Device in the Cath Lab ...
Page 68: ......
Page 74: ......
Page 75: ...References ...
Page 76: ......
Page 79: ......