22
LUCAS
®
3 Chest Compression System
Instructor Guidebook
There is a storage compartment in the middle of the Carrying Case that can hold a spare
battery and a battery charger.
The Stabilization Strap is also placed in the bag where it can be easily accessed. When
returning the Stabilization Strap to the Carrying Case, loosen the strap and roll it for the
next use.
Upper Part
The Hood
of the LUCAS device contains the Battery and the User Control Panel. Internally,
the Hood houses the compression module and all of the electronics of the LUCAS device.
Do not immerse the LUCAS device in liquid. The device can be damaged if liquid enters
the Hood.
Underneath the Hood of the LUCAS device are the Vent Holes. Do not block the Vent Holes
as this can cause the device to become too hot.
The LUCAS device has an IP43 rating, which is the level of ingress protection against
particulate matter (first number) and liquid (second number). A rating of 3 for liquid indicates
protection against spraying water, such that water falling as a spray at any angle up to 60
degrees from the vertical will not have a harmful effect. This is consistent with what you
would expect from rainfall.
Underneath the Hood is the Piston, which contains the Pressure Pad and the
Suction Cup.
Instructor Notes:
Highlight all of the components of the Upper Part of the LUCAS device. Point out each
component and demonstrate. Note: some parts, such as the control panel, may be
difficult to see from a distance. Explain that attendees will have a chance to review all of
these components up close during the hands-on portion of the training.
Instructor Notes:
After showing the Vent Holes under the Hood, point out the Pressure Pad at the end of
the Piston and Suction Cup.
Summary of Contents for LUCAS 3
Page 2: ...2 LUCAS 3 Chest Compression System Instructor Guidebook ...
Page 4: ......
Page 6: ......
Page 8: ......
Page 9: ...Chapter 1 Instructor Training Preparation ...
Page 10: ......
Page 14: ......
Page 15: ...Chapter 2 Classroom Training Introduction and Overview ...
Page 16: ......
Page 36: ......
Page 37: ...Chapter 3 Small Group Hands on Training and Application Practice ...
Page 38: ......
Page 47: ...3 SMALL GROUP HANDS ON TRAINING AND APPLICATION PRACTICE 47 ...
Page 48: ......
Page 49: ...Chapter 4 Training Evaluation ...
Page 50: ......
Page 52: ......
Page 53: ...Appendix A Important Contacts ...
Page 54: ......
Page 56: ......
Page 57: ...Appendix B Frequently Asked Questions ...
Page 58: ......
Page 66: ......
Page 67: ...Appendix C The LUCAS Device in the Cath Lab ...
Page 68: ......
Page 74: ......
Page 75: ...References ...
Page 76: ......
Page 79: ......