LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
OEM PCB/CO2 Module Cable (W26) Replacement
347
8
Removing the OEM PCB/CO2 Module Cable (W26)
Refer to
To remove the OEM PCB/CO2 module cable (
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the OEM PCB (A06) as described in
Removing the OEM PCB (A06) (p. 262)
. (This step disconnects the W26 cable from
the OEM PCB.)
4.
Remove the NIBP/CO2 module as described in
Removing the NIBP (A21)/CO2 (A23) Modules (p. 288)
.
5.
Disconnect the OEM PCB/CO2 module cable from the CO2 PCB (A23) at J4.
Installing the OEM PCB/CO2 Module Cable (W26)
Refer to
To install the OEM PCB/CO2 module cable (
1.
Connect the OEM PCB/CO2 module cable to the A23 CO2 PCB at J4.
2.
Install the NIBP/CO2 Module as described in
Installing the NIBP (A21)/CO2 (A23) Modules
3.
Install the OEM PCB (A06) as described in
4.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly
.
5.
Reassemble the case as described in
Reassembling the Case (p. 184)
.