2.3.3. Imager SETUP CD-ROM installation
The mega-pixel C-MOS sensor used in the FASTCAM-X 1280PCI system has been checked
under Photron
’
s own quality assurance criteria that sets the pixel defect allowance in each of the
geometric areas of the sensor surface. Data from each of the defective pixels on the sensor
surface is replaced with data obtained by calculating data from adjacent pixels. The defect
correction function is programmed in the attached PFV control software. Because the distribution
of defective pixels is unique to each C-MOS sensor, each of the FASTCAM-X 1280PCI must be
installed with the correction data specifically prepared for and attached to it. Use the attached
FASTCAM-X 1280PCI Imager SETUP CD-ROM for installation following the procedure below:
First, made sure the grabber board, driver and PFV have been installed.
Insert the Imager SETUP CD-ROM into the CD drive and set up automatically starts by Auto RUN.
If there is no driver or grabber board installed, the following Initialize Error message is displayed:
If PFV is not installed, you will see the following message: