24
SimplyGo User Manual
Visual, Audio
Indicators
Description
What To Do
Continuous Flow Mode
Indicator
This symbol appears when
the device is operating in
Continuous Flow Mode.
If this is your Mode
setting
, no action is
needed�
If this is not your
Mode setting
, press
the Mode button until
your proper Mode
setting appears�
1 short beep
Flow Control Setting
The large number on the
right side of the screen
shows the oxygen flow
setting� (In this example,
it is 2�) This number
can range from 1 to 6 in
increments of one-half for
Pulse and Sleep modes
and from �5 to 2 for
Continuous Flow mode.
If this is your flow
setting
, no action is
needed�
If this is not your flow
setting
, press the +
button to increase this
number or press the –
button to decrease this
number�