20
SimplyGo User Manual
Visual, Audio
Indicators
Description
What To Do
3 short beeps
+
3 long beeps
Technical Fault Alarm
This alarm occurs when:
a) the battery runs out or
b) the device has a general
malfunction and the device
is no longer operating
properly�
The device may or may not
shut down depending on
the severity of the alarm�
This alarm can be silenced
if the
symbol appears
on-screen�
Recharge the
battery and try to
restart the device�
If this alarm
occurs again,
change to another
source of oxygen
and contact
your home care
provider�
1 short beep
Low Battery Alarm
This alarm occurs when
approximately 17 minutes
of battery life remains�
Remaining battery life is
dependent on your device
settings and your activity
level� The empty battery
symbol flashes on-screen�
Replace the
battery or connect
to a power source�
(Press the Alarm
symbol to silence
the alarm�)
3 long beeps
+
1 short beep
No Flow Alarm
This alarm occurs when
the device detects there is
no oxygen flowing in the
patient cannula�
Check the cannula
for kinks or other
obstructions that
are stopping
oxygen flow
through it�