Chapter 6 - Disassembly/Reassembly Guide
37
6
Disassembly/Reassembly Guide
Warning
Performance verification: do not place the monitor into operation after
repair or maintenance has been performed, until all performance tests and
safety tests listed in Chapter 8 of this service manual have been performed.
Failure to perform all tests could result in erroneous monitor readings, or
patient/operator injury.
Warning
Energized circuits are accessible with the covers open. Do not work on the
monitor with the covers open. Only qualified service personnel should
open or disassemble the monitor.
Introduction
Warning
Before attempting to open or disassemble the monitor, disconnect the
power supply from the monitor.
Warning
High voltage is generated by the LCD backlight driver. Do not operate the
monitor with the covers open.
Summary of Contents for FM-2