Chapter 8 - Test and Inspection Matrix
77
8
Test and Inspection Matrix
Warnings
Warning
EXPLOSION HAZARD! Do not use the FM-2 in a flammable atmosphere
where concentrations of flammable anesthetics or other materials may
occur.
Warning
SHOCK HAZARD! The power-receptacle must be a three-wire grounded
outlet. A hospital grade outlet is required. Never adapt the three-prong plug
from the power supply or accessory to fit a two-slot outlet. If the outlet has
only two slots, make sure that it is replaced with a three-slot grounded
outlet before attempting to operate the monitor.
Warning
SHOCK HAZARD! Do not attempt to connect or disconnect a power cord
with wet hands. Make certain that your hands are clean and dry before
touching a power cord.
Warning
Use only patient cables and transducers supplied with the monitor. Use of
any other patient cables may result in out-of-specification performance and
possible safety hazards.