Safety Testing
Chapter 8 - Test and Inspection Matrix
87
Test Steps
1. Connect the external power supply output cord to the monitor. Then
connect the appropriate external power supply input power cord to the
analyzer as recommended by the analyzer operating instructions.
2. Using the appropriate test cable, connect the analyzer to a fixing nut of
the RS-232 connector on the back of the monitor.
3. Turn the monitor on.
4. Perform the test as recommended by the analyzer operating instructions.
The analyzer leakage current indication must not exceed the values listed in
Table 22.
Table 22
Enclosure Leakage Current
Test Condition
Polarity
Maximum Leakage Current
(µA)
IEC 601-1 ES1
AAMI/ANSI
Standard
Normal
Normal
100 µA
100 µA
Reversed
100 µA
100 µA
S.F.C.
1
(Open Supply)
1. S.F.C. = Single Fault Condition
Normal
500 µA
300 µA
Reversed
500 µA
300 µA