Safety Testing
86
Chapter 8 - Test and Inspection Matrix
Ground Integrity
The FM-2 is a Class 2, Type BF instrument. It has an external power supply, and
is constructed with double insulation. The FM-2
does not require an isolated
Earth Ground terminal, neither is one installed. No Protective Ground
Continuity check is required.
Electrical Leakage
The following tests verify the electrical leakage of the monitor:
• Enclosure leakage Current
• Patient Leakage Current
Safety Test 1: Enclosure Leakage Current Test
Normal
Condition
This test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied
Parts. The test measures leakage current of exposed metal parts of the
Instrument Under Test. It tests both normal and reversed polarity.
For type BF and CF Applied Parts the test measures AP/GND.
Safety Test according IEC 601-1 (Clause 19).
Report the highest value.
Single Fault
Condition
(S.F.C.) Open
Supply
This test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied
Parts. The test measures leakage current of exposed metal parts of the
Instrument Under Test with one supply lead interrupted. It tests both normal
and reversed polarity.
For type BF and CF Applied Parts the test measures AP/GND.
Safety Test according IEC 601-1 (Clause 19).
Report the highest value.
Summary of Contents for FM-2
Page 4: ......
Page 10: ......
Page 12: ......
Page 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Page 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Page 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...