background image

-

The blood pressure monitor has 2 user
profiles. If both user profiles are in use,
choose an existing profile to overwrite.

-

You can also delete both user profiles by
pressing and holding the user ID button for
approx. 10 seconds. The display of the
device shows 'del'. All stored data is deleted
and you have to follow steps 1-4 to pair and
add a new user. 

5 The blood pressure monitor shows the

Bluetooth icon on the display as soon the
connection has been established and switches
off automatically after a few seconds. 

When the blood pressure monitor is successfully
paired with your mobile device, the blood pressure
monitor automatically transmits your personal
health data to your mobile device via Bluetooth

®

Smart.
Note: Only when the Philips HealthSuite health
app is active, your personal health data can be
transmitted.

Measuring blood pressure

Tips for proper measurement

- Rest for 5 minutes before you measure your

blood pressure.

- Wait at least 3 minutes between measurements.

This allows your blood circulation to recover.

- For a meaningful comparison, try to measure

under similar conditions. For example, take
daily measurements at approximately the same
time, on the same arm, or as directed by your
physician.

- For a good Bluetooth connection between the

blood pressure monitor and your mobile
device, make sure the two are close and there
are no obstacles between the two devices. We
recommend not to have the two devices farther
than 16 feet (5 meters) apart.

We advise you not to take a measurement under
the following circumstances, as this measurement
may not be representative:
- Within 1 hour after eating or drinking
- Immediately after smoking
- While bathing and within 20 minutes after

taking a bath

- While you are talking or moving your arm, hand

or fingers

- In a very cold environment
- When you need to urinate 

Attaching the cuff

1

Remove all jewelry, such as watches and
bracelets from your left arm.
Note: If your physician has diagnosed you with
poor circulation in your left arm, use your right
arm.

2 Roll or push up your sleeve to expose the

skin. Make sure your sleeve is not too tight. 

3 Hold your arm with your palm facing up and

slide the cuff onto your left upper arm (Fig. 10).

4 Position the lower edge of the cuff 

¾

" - 1

¼

" (2-3

cm) above the crease of the elbow.

5 Fasten the cuff around your arm, leaving no

extra room between the cuff and your skin. If
the cuff is too loose, the measurement will not
be accurate.

-

The cuff will not cause any potential
sensitization or irritation of the skin. The
materials of the cuff have been tested and
found to comply with requirements of ISO
10993-5:2009, ISO 10993-1:2009 and ISO
10993-10:2010. 

6 Position yourself in the correct way for proper

measurement:

-

Make sure you do not wear tight clothing
during measurement.

-

Sit comfortably with legs uncrossed, feet flat
on the floor. Make sure that you sit upright
with your back straight.

-

The center of the cuff should be at the same
level as the heart.

-

Relax your wrist and hand. Do not bend your
wrist back, clench your fist, or bend your
wrist forward.

Start measurement

1

Press the user ID button (Fig. 11) or 'on' button
(Fig. 12) once, to switch on the device. The
device automatically selects the previous user.

-

To change the user profile, press the user
ID button (Fig. 11) and select a different user
(Fig. 13). Make sure the correct user is
selected, so the measurement data is
properly transmitted and stored. It is not
possible to switch a user profile after a
measurement.

-

When the health app is open, the app
automatically selects the correct user profile.
In this case, the user profile can be changed
by either closing the app and reopening it
again with the correct user profile, or by
closing the app and using the user ID button.

-

Also a guest user can be selected. A guest
user is a user without a user profile in the
app. The guest user is for performing a
measurement on other people without a
user profile in the health app. Measurements
performed when using the guest user are
not stored in the memory nor transmitted to
the app.

2 Attach the cuff to your arm (see 'Attaching the

cuff') and make sure your posture is correct (see
'Tips for proper measurement').

3 Press the 'on' button to start the measurement

(Fig. 12). All display characters are briefly shown
on the display (Fig. 14). The device is ready for
measurement and the number 0 appears (Fig.
15). Inflation of the cuff starts automatically.

-

During inflation, the unit determines the
systolic pressure and diastolic pressure as
well as heart rate. This is shown by the
heart rate detection symbol. 

-

The movement detector will light up when
movement is detected. This may result in
inaccurate measurement results. 

Summary of Contents for DL8760

Page 1: ...Koninklijke Philips N V All rights reserved Manufactured for Philips Consumer Lifestyle A division of Philips Electronics North America Corporation P O Box 10313 Stamford CT 06904 4222 100 5747 1 1 2016 1 2 2 3 4 5 6 7 8 9 2 3 cm 1 10 11 12 13 14 15 16 17 18 ...

Page 2: ...plugged in Consult your physician if you suffer from illnesses prior to using the device No modifications of this equipment are allowed This may result in increased emissions or decreased immunity of the device Do not use while bathing and within 20 minutes after taking a bath Do not use the blood pressure monitor during charging as this can cause injury Do not touch the output of the DC charger a...

Page 3: ...er the tap This device is not suitable for continuous monitoring during medical emergencies or operations This device cannot be used with HF High Frequency surgical equipment at the same time Do not use the DC charger in or near a power outlet that contains an electric air freshener to prevent damage to the DC charger Keep the device away from fire and heat sources as the battery can overheat caus...

Page 4: ... symbol The device uses Bluetooth for communication Heart rate detection Heart rate detection during the measurement Battery status indications Battery symbol Battery status The battery is almost empty The battery is empty When you measure 3 times a day starting with a fully charged battery the device can be used for about 20 days until a recharge is needed In case of normal use the battery can be...

Page 5: ...rtbeat intervals and calculates the standard deviation If the standard deviation exceeds a pre defined threshold the irregular heart rate detector symbol lights up when the measurement results are displayed Fig 5 Caution The appearance of the irregular heart rate detector symbol indicates that a heart rate irregularity was detected during measurement Usually this is not a cause for concern Due to ...

Page 6: ... the cuff 1 2 3 cm above the crease of the elbow 5 Fasten the cuff around your arm leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate The cuff will not cause any potential sensitization or irritation of the skin The materials of the cuff have been tested and found to comply with requirements of ISO 10993 5 2009 ISO 10993 1 2009 and IS...

Page 7: ... only be done for that user Cleaning and storage Caution This device is not washable Never immerse the device in water and do not rinse it under running water Caution Avoid sudden movements and hard contacts with objects Caution Never use compressed air scouring pads abrasive cleaning agents or aggressive liquids such as petrol or acetone to clean the device 1 Switch off the device and unplug the ...

Page 8: ...ifferent during each measurement For a meaningful comparison try to measure under similar conditions For example take measurements daily at approximately the same time on the same arm or as directed by a physician Fluctuations of blood pressure during the day are normal Blood pressure fluctuates from minute to minute and normally shows a circadian rhythm over a 24 hour period with highest readings...

Page 9: ...ct the correct user profile on the blood pressure monitor before your measurement Otherwise the data cannot be transmitted to your app Repeat the measurement with the correct profile selected Specifications Product name Philips upper arm blood pressure monitor with Bluetooth Smart Power supply 3 7V 1000mAH built in rechargeable li polymer battery 6V 1A DC charger Display Display with white LED bac...

Page 10: ...vice means protected against access to hazardous parts with a finger and against vertically falling water drops when tilted up to 15 degrees 4ºF 140ºF Indicates the storage and transportation temperature limits to which the medical device can be safely exposed 4 F to 140 F 20 C to 60 C Indicates the relative humidity limits to which the device can be safely exposed 10 to 93 Symbol for the 2 year P...

Page 11: ...cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s Electrical power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power fre quency 50 60 Hz magnet ic fi...

Page 12: ...78 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separation distance for the hi...

Page 13: ... equipment For assistance visit our website www philips com support or call toll free 1 844 531 6861 BlueTooth wordmark The BlueTooth Smart wordmark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Philips is under license App Store and iPhone App Store and iPhone are trademarks of Apple Inc registered in the U S and other countries App Store is a service...

Reviews: