Problem

Possible cause

Solution

The display

shows Err

Communication

error.

Check if the app is

on and try data

transmission

again.

The display

shows E3

The cuff is not

properly

secured.

Refasten the cuff,

wait 3 minutes

and then measure

again.

The display

shows E10

or E11

The device

detected

motion, talking

or the heart

rate is too weak

during the

measurement.

Wait for 3

minutes and then

measure again.

Do not move

during

measurement. 

The display

shows E20

The device

does not detect

the heart rate

signal.

Make sure the

device is in

contact with the

skin. Loosen the

clothing on the

arm and measure

again.

The display

shows E21

The

measurement

failed.

Wait for 3 minutes

and then measure

again.

The display

shows EExx

A system error

occurred.

Retake the

measurement. If

the problem

persists, call

1-844-531-6861

for assistance.

Data

transmis­

sion or

pairing

failed.

Bluetooth is off. Turn on Bluetooth

on your mobile

device.

The Philips

HealthSuite

health app is

off.

Press the icon on

your mobile

device to activate

the health app.

The blood

pressure

monitor and

mobile device

are more than

16 feet (5

meters) feet

away from

each other.

Place your mobile

device closer to

the blood

pressure monitor.

You selected

the wrong

profile on the

blood pressure

monitor.

Select the correct

user profile on the

blood pressure

monitor before

your

measurement.

Otherwise the

data cannot be

transmitted to

your app. Repeat

the measurement

with the correct

profile selected

Specifications

Product name

Philips upper arm blood

pressure monitor with

Bluetooth® Smart 

Power supply

3.7V 1000mAH built-in

rechargeable li-polymer

battery, 6V 1A DC charger

Display 

Display with white

LED backlight

Visible area = 3 3/8" (L) x

15/16" (W) / 86.1 mm (L) x

24 mm (W)

Measurement method Oscillometric method

Measurement range

Rated cuff pressure:

0mmHg - 300mmHg.

Measurement pressure:

40mmHg - 230mmHg

Heart rate: 40-199 beats

per minute

Accuracy

Pressure: 41°F to 104°F /

5°C - 40°C within ±

3mmHg. Heart rate: ±5%

of measurement result

on display

Normal operating

condition

Temperature: 41°F to

104°F/ 5°C to 40°C.

Relative humidity:

85%RH. Atmospheric

pressure: 86kPa to

106kPa

Storage and

transportation

conditions

Temperature: -4°F to

140°F / -20°C to 60°C.

Relative humidity: 10% to

93%. Atmospheric

pressure: 50kPa to

106kPa

Measurement

perimeter of the

upper arm

About 8 

¾

 - 16 

½

 inch

(22- 42 cm)

Net weight

Approx. 9.34 oz / 265g

External dimensions

Approx. 5 1/8" × 2 7/8" × 1

3/16"

(130.9mm × 73mm ×

29.4mm)

Accessories

DC charger, user manual

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against

ingress of water

IP22, This means:

protected against access

to hazardous parts with a

finger and against

vertically falling water

drops when tilted up to

15 degrees.

Summary of Contents for DL8760

Page 1: ...Koninklijke Philips N V All rights reserved Manufactured for Philips Consumer Lifestyle A division of Philips Electronics North America Corporation P O Box 10313 Stamford CT 06904 4222 100 5747 1 1 2016 1 2 2 3 4 5 6 7 8 9 2 3 cm 1 10 11 12 13 14 15 16 17 18 ...

Page 2: ...plugged in Consult your physician if you suffer from illnesses prior to using the device No modifications of this equipment are allowed This may result in increased emissions or decreased immunity of the device Do not use while bathing and within 20 minutes after taking a bath Do not use the blood pressure monitor during charging as this can cause injury Do not touch the output of the DC charger a...

Page 3: ...er the tap This device is not suitable for continuous monitoring during medical emergencies or operations This device cannot be used with HF High Frequency surgical equipment at the same time Do not use the DC charger in or near a power outlet that contains an electric air freshener to prevent damage to the DC charger Keep the device away from fire and heat sources as the battery can overheat caus...

Page 4: ... symbol The device uses Bluetooth for communication Heart rate detection Heart rate detection during the measurement Battery status indications Battery symbol Battery status The battery is almost empty The battery is empty When you measure 3 times a day starting with a fully charged battery the device can be used for about 20 days until a recharge is needed In case of normal use the battery can be...

Page 5: ...rtbeat intervals and calculates the standard deviation If the standard deviation exceeds a pre defined threshold the irregular heart rate detector symbol lights up when the measurement results are displayed Fig 5 Caution The appearance of the irregular heart rate detector symbol indicates that a heart rate irregularity was detected during measurement Usually this is not a cause for concern Due to ...

Page 6: ... the cuff 1 2 3 cm above the crease of the elbow 5 Fasten the cuff around your arm leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate The cuff will not cause any potential sensitization or irritation of the skin The materials of the cuff have been tested and found to comply with requirements of ISO 10993 5 2009 ISO 10993 1 2009 and IS...

Page 7: ... only be done for that user Cleaning and storage Caution This device is not washable Never immerse the device in water and do not rinse it under running water Caution Avoid sudden movements and hard contacts with objects Caution Never use compressed air scouring pads abrasive cleaning agents or aggressive liquids such as petrol or acetone to clean the device 1 Switch off the device and unplug the ...

Page 8: ...ifferent during each measurement For a meaningful comparison try to measure under similar conditions For example take measurements daily at approximately the same time on the same arm or as directed by a physician Fluctuations of blood pressure during the day are normal Blood pressure fluctuates from minute to minute and normally shows a circadian rhythm over a 24 hour period with highest readings...

Page 9: ...ct the correct user profile on the blood pressure monitor before your measurement Otherwise the data cannot be transmitted to your app Repeat the measurement with the correct profile selected Specifications Product name Philips upper arm blood pressure monitor with Bluetooth Smart Power supply 3 7V 1000mAH built in rechargeable li polymer battery 6V 1A DC charger Display Display with white LED bac...

Page 10: ...vice means protected against access to hazardous parts with a finger and against vertically falling water drops when tilted up to 15 degrees 4ºF 140ºF Indicates the storage and transportation temperature limits to which the medical device can be safely exposed 4 F to 140 F 20 C to 60 C Indicates the relative humidity limits to which the device can be safely exposed 10 to 93 Symbol for the 2 year P...

Page 11: ...cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s Electrical power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power fre quency 50 60 Hz magnet ic fi...

Page 12: ...78 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separation distance for the hi...

Page 13: ... equipment For assistance visit our website www philips com support or call toll free 1 844 531 6861 BlueTooth wordmark The BlueTooth Smart wordmark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Philips is under license App Store and iPhone App Store and iPhone are trademarks of Apple Inc registered in the U S and other countries App Store is a service...

Reviews: