background image

pressure measurements, heart rate (pulse)
measurements, transmit data via Bluetooth®
Smart to your mobile device and display your
personal measurement results in the Philips
HealthSuite health app. The device can also be
used as a standalone device.
This user manual contains important safety
information and provides step-by-step instructions
for using the blood pressure monitor.
Read this information carefully before you use the
device and save it for future reference.
Features:
- 3 7/16 in x 15/16 in / 86.5mm×24mm display

with white backlight

- Measure-during-inflation technology
- Supports 2 users

Intended use

The Philips upper arm blood pressure monitor is a
digital monitor intended for use in measuring blood
pressure and heart rate in adult patient population
with arm circumference ranging from 8 

¾

 - 16 

½

inch ( 22cm to 42cm ). The device is intended to be
used in an indoor home environment. 

Display

 

ym

bol

Description

Explanantion

Systolic

blood

pressure

Maximum blood pressure,

see also section systolic and

diastolic pressure.

Diastolic

blood

pressure

Minimum blood pressure,

see also section systolic and

diastolic pressure.

Heart rate

Number of heartbeats per

minute (pulse is typically

equivalent to heart rate).

Battery status Indicates status of battery

during charging.

Measurement

unit

Measurement unit of blood

pressure.

Irregular heart

rate detector

Irregular heart rate detection

during the measurement.

User IDs

Start measurement for

selected user, and transmit

the measurement result.

Movement

detector

Moving during the

measurement will result in

an inaccurate result.

Blood

pressure

classification

Classification of measured

blood pressure following

WHO system (see 'Blood

pressure classification').

Bluetooth®

symbol

The device uses Bluetooth

for communication.

Heart rate

detection

Heart rate detection during

the measurement.

Battery status indications

Battery

symbol

Battery status

The battery is almost empty.

+

The battery is empty.

When you measure 3 times a day starting with a
fully charged battery, the device can be used for
about 20 days until a recharge is needed. In case
of normal use, the battery can be charged around
300 times.
Note: Data will be lost when the battery is
completely empty. 

Charging

The battery of this device is a built-in rechargeable
li-polymer battery with a capacity of 1000 mAh.
Use the original DC charger supplied to charge the
battery. The DC charger transforms 100-240V AC
to 6V DC.
When the battery is empty, it takes approx. 2 hours
to fully charge the battery of the device.
1

Put the small DC plug in the socket of the
device (Fig. 2).

2 Put the DC charger in the electrical outlet.
To ensure an optimal life time of the battery in the
product it is recommended to store it 50% charged
and re-charge every 3 months.

Battery charging indications

Battery

symbol

Battery charging indication

Battery charging: half full

Battery charging: almost full

Battery fully charged

Using the blood pressure monitor

This tubeless device uses the oscillometric method
to measure blood pressure and heart rate. 
Before every measurement, the unit establishes a
“zero point” equivalent to the atmospheric
pressure. Then it starts inflating the cuff. During the
measurement, the device detects the pressure
oscillations in the blood vessels generated by the
heart pumping blood through the body. These
pressure oscillations are used to determine systolic
and diastolic blood pressure as well as heart rate.
While measuring heart rate, the device also
determines the small variations between the
individual heartbeats. If these variations exceed a
pre-defined threshold, the irregular heart rate
detector symbol lights up.

Summary of Contents for DL8760

Page 1: ...Koninklijke Philips N V All rights reserved Manufactured for Philips Consumer Lifestyle A division of Philips Electronics North America Corporation P O Box 10313 Stamford CT 06904 4222 100 5747 1 1 2016 1 2 2 3 4 5 6 7 8 9 2 3 cm 1 10 11 12 13 14 15 16 17 18 ...

Page 2: ...plugged in Consult your physician if you suffer from illnesses prior to using the device No modifications of this equipment are allowed This may result in increased emissions or decreased immunity of the device Do not use while bathing and within 20 minutes after taking a bath Do not use the blood pressure monitor during charging as this can cause injury Do not touch the output of the DC charger a...

Page 3: ...er the tap This device is not suitable for continuous monitoring during medical emergencies or operations This device cannot be used with HF High Frequency surgical equipment at the same time Do not use the DC charger in or near a power outlet that contains an electric air freshener to prevent damage to the DC charger Keep the device away from fire and heat sources as the battery can overheat caus...

Page 4: ... symbol The device uses Bluetooth for communication Heart rate detection Heart rate detection during the measurement Battery status indications Battery symbol Battery status The battery is almost empty The battery is empty When you measure 3 times a day starting with a fully charged battery the device can be used for about 20 days until a recharge is needed In case of normal use the battery can be...

Page 5: ...rtbeat intervals and calculates the standard deviation If the standard deviation exceeds a pre defined threshold the irregular heart rate detector symbol lights up when the measurement results are displayed Fig 5 Caution The appearance of the irregular heart rate detector symbol indicates that a heart rate irregularity was detected during measurement Usually this is not a cause for concern Due to ...

Page 6: ... the cuff 1 2 3 cm above the crease of the elbow 5 Fasten the cuff around your arm leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate The cuff will not cause any potential sensitization or irritation of the skin The materials of the cuff have been tested and found to comply with requirements of ISO 10993 5 2009 ISO 10993 1 2009 and IS...

Page 7: ... only be done for that user Cleaning and storage Caution This device is not washable Never immerse the device in water and do not rinse it under running water Caution Avoid sudden movements and hard contacts with objects Caution Never use compressed air scouring pads abrasive cleaning agents or aggressive liquids such as petrol or acetone to clean the device 1 Switch off the device and unplug the ...

Page 8: ...ifferent during each measurement For a meaningful comparison try to measure under similar conditions For example take measurements daily at approximately the same time on the same arm or as directed by a physician Fluctuations of blood pressure during the day are normal Blood pressure fluctuates from minute to minute and normally shows a circadian rhythm over a 24 hour period with highest readings...

Page 9: ...ct the correct user profile on the blood pressure monitor before your measurement Otherwise the data cannot be transmitted to your app Repeat the measurement with the correct profile selected Specifications Product name Philips upper arm blood pressure monitor with Bluetooth Smart Power supply 3 7V 1000mAH built in rechargeable li polymer battery 6V 1A DC charger Display Display with white LED bac...

Page 10: ...vice means protected against access to hazardous parts with a finger and against vertically falling water drops when tilted up to 15 degrees 4ºF 140ºF Indicates the storage and transportation temperature limits to which the medical device can be safely exposed 4 F to 140 F 20 C to 60 C Indicates the relative humidity limits to which the device can be safely exposed 10 to 93 Symbol for the 2 year P...

Page 11: ...cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s Electrical power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power fre quency 50 60 Hz magnet ic fi...

Page 12: ...78 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separation distance for the hi...

Page 13: ... equipment For assistance visit our website www philips com support or call toll free 1 844 531 6861 BlueTooth wordmark The BlueTooth Smart wordmark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Philips is under license App Store and iPhone App Store and iPhone are trademarks of Apple Inc registered in the U S and other countries App Store is a service...

Reviews: