Receiver VCXO Test
Chapter 8 - RF Bench Repair
Figure 23 Receiver VCXO Test
The Spectrum Analyser should be set as follows:
START frequency
400 MHz
STOP frequency
500 MHz
Resolution Bandwidth
100 kHz
Video Bandwidth
100 kHz
Reverence Level
+10 dBm
The Spectrum Analyser measurement must be:
Frequency
Channel Fre 21 .400 MHz
+/- 5kHz
Level
> - 6 dBm
2. If the level is lower the Receiver VCXO is to be replaced. If the
frequency and level are correct the Receiver Assembly needs to be
replaced.