Chapter 4 - Troubleshooting
FMP
Figure 9 Troubleshooting: FMP
FMP Test
To test an ultrasound transducer:
1. Connect the telemetry receiver to the fetal monitor using the interface cable.
2. Switch on the fetal monitor, its recorder and the telemetry receiver.
3. Connect the transducer to the ultrasound socket.
4. Switch on the transmitter.
5. Increase the loudspeaker volume on the fetal monitor to an audible level.