Chapter 4 - Troubleshooting
4
Troubleshooting
Solving General Problems
Troubleshooting is based on the interpretation of the Receiver LED’s. This
chapter describes common problems and their causes, and provides
troubleshooting flowcharts and guidance for their solution.
Problem
Possible Causes
Solutions
All the lights on the
receiver stay on when
the receiver is turned
on.
Fetal monitor is switched off.
Possible defect in the receiver, interface
cable or fetal monitor.
Switch on fetal monitor.
Carry out the tests described in “Testing the
Receiver” on page 11.
The Telemetry
Indicator Lamp on the
fetal monitor does not
light when the monitor
and the receiver are
switched on.
Incorrect interface connection between
the monitor and the receiver.
Faulty interface cable.
Follow the instructions in Instructions for Use for
details on how to connect the monitor to the
receiver.
Replace interface cable.
Receiver Power On
Light does not light
when the receiver is
switched on.
Power cable not plugged into the power
supply.
Fuses need replacing.
Plug in and switch on.
Replace fuses.
Transmission INOP
light on the receiver is
still lit when the
transmitter is switched
on.
Receiver and transmitter do not have
the same channel or serial number.
Batteries in the transmitter are
exhausted.
Check channel number and the serial numbers
are the same on the receiver and the transmitter.
Change the batteries in the transmitter (refer to
the Instructions for Use for details.)
Battery Low Light lit on
receiver.
Power in batteries is low.
Change batteries.
INOP transmission
lamp is lit after the
patient has moved a
short distance away
form the receiver.
Local Antenna:
Antenna not connected correctly.
Remote Antenna System:
Antenna cable not connected correctly
to receiver.
Check antenna is connecting correctly.
Test the antenna system by bringing the
transmitter close to the receiver. If the
transmission is good, then the antenna system is
not functioning properly.
Summary of Contents for 50 T Series
Page 4: ......
Page 10: ...x List of Figures ...
Page 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Page 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Page 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Page 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Page 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Page 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Page 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Page 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Page 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...