Power Supply Board
Chapter 7 - Replacing Parts
Power Supply Board
To remove the Power Supply Board, first, follwing the instructions for
dismantling the Receiver, then:
1. Remove power switch manipulator (A).
2. Remove flat cable from processor board (B).
3. Remove 7 screws (C).
4. Remove ground cable. Keep the ground cable screw-washer in a safe
place. (D)
5. Lift clear of the housing.
Removing the Power Supply Board
To replace the Power Supply Board:
1. Put power supply board into place.
2. Attach the power supply board with the 8 screws.