Section 9: Technical Specifications
Environmental
Operating
Storage
Temperature
5° C to 40° C (41° F to 104° F)
-20° C to 60° C (-4° F to 140° F)
Relative Humidity
15 to 95% (non-condensing)
15 to 95% (non-condensing)
Atmospheric Pressure
101 kPa to 77 kPa
(approximately 0-2286 m/0-7500 ft)
n/a
Physical
Dimensions:
22.23 cm W x 18.42 cm L x 10.80 cm H
(8.75” W x 7.25” L x 4.25” H)
Weight:
Approximately 2 kg (4.4 lbs)
Standards Compliance
This device is designed to conform to the following standards:
• IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance
• IEC 60601-1-2: Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance –
Collateral standard: Electromagnetic compatibility - Requirements and tests
• IEC 60601-1-6: Medical Electrical Equipment – Part 1-6: General Requirements for Safety and Essential Performance – Collateral
standard: Usability
• IEC 60601-1-8: Medical Electrical Equipment – Part 1-8: General Requirements for Basic Safety and Essential Performance – Collateral
standard: General Requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
• IEC 60601-1-11: Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral
standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
• ISO 10651-6: Lung Ventilators for Medical Use – Particular Requirements for Basic Safety and Essential Performance, Part 6. Home
care ventilatory support devices
• RTCA DO-160F Section 21, Category M; Emission of Radio Frequency Energy
Electrical
AC Voltage Source:
100 to 240 VAC, 50/60 Hz, 1.2 A
DC Power Source:
12 VDC, 5.0 A (External Battery)
24 VDC, 4.2 A (Power Supply)
Type of Protection Against Electric Shock:
Class II (To be used with external Class II power supply only)
Degree of Protection Against Electric Shock:
Type BF Applied Part
Degree of Protection against Ingress of Water:
Drip Proof, IP22
Mode of Operation:
Continuous
SD Card and SD Card Reader
Use only SD cards and SD card readers available from Philips Respironics, including the following:
SanDisk ® Card Reader/Writer - SanDisk ImageMate - REF SDDR-99-A15