35
Alarm
Priority
Audible
Visual Indicators
Device Action
User Action
Low Battery
(When Battery
is Attached)
Escalates
from
Medium to
High
◊◊◊
(Medium -
when approx.
20 minutes
remains)
◊◊◊ ◊◊
(High - when
approx. 10
minutes
remains)
Medium Priority -
Yellow flashing button;
Low External Battery
message appears in
yellow; on status panel,
box around battery is
yellow.
High Priority - Red
flashing button; Low
External Battery
message appears in
red; on status panel,
box around battery
is red.
Operates
Switch to an alternate battery, or
switch to AC power.
AC Power
Disconnected
(When Battery
is Attached)
Medium
◊◊◊
Yellow flashing
button; AC Power
Disconnected
message, and a box
appears around
battery in use.
Switches to
alternate
power source
Check AC power adapter and
reconnect it if it has become
disconnected. Make sure device is
not connected to an overloaded
AC circuit.
Stuck Key
Low
◊◊
Solid yellow button;
Keypad Stuck message.
Operates
Check the keys to determine if
they are lodged in the case. If alarm
continues, place patient on alternate
source of therapy and contact
product support.
Insert SD
Card (When
Oximeter is
Attached)
Low
◊◊
Solid yellow button;
Insert SD Card
message
Operates
Insert an SD card into the device or
remove the oximeter.
Informational Messages
Message
Priority
Audible
Visual Indicators
Device Action
User Action
Card Error
Info
◊
Card Error message
Operates
Remove SD Card and use another
card, if available. Ensure card meets
specifications. If condition persists,
contact product support.
Heated Tube
Failure
Info
None
Flashing icon:
Device
operate;
Humidifier
shuts down
Tubing may be overheating or
malfunctioning. Turn off airflow
and reconnect the heated tubing
to the humidifier according to the
humidifier instructions. If the alert
continues to occur, contact product
support.
Humidifier
Failure
Info
None
Flashing icon:
Device
operate;
Humidifier
shuts down
Humidifier failure. Turn off airflow
and reconnect the humidifier to the
device according to the humidifier
instructions. If the alert continues to
occur, contact product support.