23
Mode
Setting
CPAP
Auto-CPAP
Flex Type
C-Flex, C-Flex+
C-Flex, A-Flex
Flex
Off, 1, 2, 3
Off, 1, 2, 3
CPAP
4-20 cm H
2
O
(0.5 cm H
2
O res)
n/a
Maximum Pressure
n/a
Min Pressure-20 cm H
2
O
(0.5 cm H
2
O res)
Minimum Pressure
n/a
4-Max Pressure cm H
2
O
(0.5 cm H
2
O res)
Ramp Time
0-45 min
(5 min res)
0-45 min
(5 min res)
Ramp Start Pressure
4-(CPAP-0.5) cm H
2
O
(0.5 cm H
2
O res)
4-Min Pressure cm H
2
O
(0.5 cm H
2
O res)
The following list gives more detail on the Settings and Alarms listed in the previous tables.
Mode
Change the Mode setting to one of the following therapy modes:
• CPAP
• T
• Auto-BiLevel
• S
• PC
• autoSV
• S/T
• Auto-CPAP
• AVAPS-AE
Flex Type
This allows you to choose the Flex comfort feature setting.
Flex
Set Flex to 1, 2, or 3 to enable the setting. A setting of 1 provides a small amount of pressure relief, with higher
numbers providing additional relief. Select Off to disable the setting.
AVAPS
Select On or Off to enable or disable AVAPS.
CPAP
Increase or decrease the CPAP pressure setting from 4 to 20 cm H
2
O in increments of 0.5.
IPAP
Increase or decrease the Inspiratory Positive Airway Pressure (IPAP) from 4 to 30 cm H
2
O in increments of
0.5. You cannot set the IPAP setting lower than the EPAP setting. IPAP is limited to 25 cm H
2
O when the Flex
feature is enabled.
IPAP Max Pressure
Increase or decrease the setting from 4 to 30 cm H
2
O in increments of 0.5. The IPAP Max Pressure must be
equal to or greater than the IPAP Min value.
IPAP Min Pressure
Increase or decrease the setting from 4 to 30 cm H
2
O in increments of 0.5. The IPAP Min Pressure must be
equal to or greater than the EPAP value, and it must be less than or equal to the IPAP Max Pressure.