6CP30/15-3
2-2
CONTROLS AND INDICATORS
Control Panel (Cont’d)
4.
High Voltage Off/Ready.
Momentary switch turns high voltage off when activated.
Ready
lamp
illuminates when all pre-conditions required for high voltage activation are met.
Conditions required for
Ready
lamp illumination are:
Power applied to unit and
Main Power Breaker
(1) on,
Main Power
lamp illuminated (2).
External Interlock
circuit series loop must not be open (10).
Voltage Control
(5) must be at zero. (Zero Start.)
Overcurrent circuit must be reset if
Reset
lamp is illuminated (7).
5.
Voltage Control.
Turn clockwise to raise output voltage. High voltage cannot be activated if
Voltage
Control
is not started at zero.
6.
Overcurrent Trip Adjust.
Dial adjusts from 1 to 11 corresponding to 10% to 110% of rated output
current.
Overcurrent Trip/Reset
lamp illuminates and high voltage turns off when output current exceeds
setting, causing circuit to trip. Circuit also acts as short circuit and overload protection on high voltage
output. To reactivate high voltage,
Voltage Control
(5) must be returned to zero, and
Overcurrent Reset
switch (7) must be pressed to clear overcurrent circuit.
7.
Overcurrent Reset.
Reset
lamp illuminates to show that overcurrent circuit has tripped. High voltage
circuits are deactivated. Momentary
Reset
switch must be pressed to extinguish
Reset
lamp to allow high
voltage to be reapplied after returning
Voltage Control
(5) to zero.
8.
Currentmeter.
Displays output current of test set, units based on current meter range setting.
9.
Voltmeter.
Displays output voltage of test set, units in kilovolts.
10.
Ext.
Intlk.
– SX1.
Provides for user connection of external interlock or auxiliary safety control device such
as emergency off switch, gate switch, foot switch, dead man switch, etc. Included plug has shorting
jumper installed to complete circuit. Jumper must be removed and cable connected to user supplied
device by user if desired to use this provision. Only non-energized switch or dry relay contact devices may
be used. This is a series 120VAC circuit that must remain closed for high voltage to be activated.
11.
TX2.
Provided interconnect cable is connected here and to the same designation connector on the high
voltage unit. This is the metering and return connection.
Do not attempt to operate high voltage unit
without this cable connected at both ends.
12.
TX1.
Provided power interconnect cable connects here and to the same designation connector on high
voltage unit.
13.
Line Cord.
Supplies input power to unit.