SECTION 5: CALIBRATION
CAUTION: Calibration should only be done by persons familiar with High Voltage
testing and safety procedures.
All calibrations have been done at the factory. Periodic calibration of the output voltmeter and output
currentmeter should be done annually.
NOTE: Refer to Electrical Diagram Section for schematic pertaining to the model number of your test set.
Locating the Calibration Adjustments
The calibration points are shown in the following diagram.
1.
Output Voltmeter
Voltmeter is peak responding calibrated to Peak/√2 (RMS value of sine wave).
15kV Range:
Connect a precision high voltage voltmeter between the 15kV output (side of tank) to
ground. Be sure the
Measurement Divider
switch on the control panel is in the 15kV
position. Raise the output to approximately 80% of the rated output voltage. Adjust the
reading on the panel meter (M2) by means of potentiometer R1 to a corresponding
reading. Check linearity at several points from 20% to 100% of rated tap voltage.
30kV Range:
Connect a precision high voltage voltmeter between the 30kV output (top of tank) to
ground. Be sure the
Measurement Divider
switch on the control panel is in the 30kV
position. Raise the output to approximately 80% of the rated output voltage. Adjust the
reading on the panel meter (M2) by means of potentiometer R2 to a corresponding
reading. Check linearity at several points from 20% to 100% of rated tap voltage.
R101
10%
Ovld
R103
110%
Ovld
R235
Range
Ovld
R201
200mA
CM
R202
20mA
CM
R203
2mA
CM
R204
200µA
CM
R2
30 kV
VM
R1
15 kV
VM
NOTE:
R205, R206, and R305, R306 are set at the factory and should not be adjusted.
R205
R206
R305
R306