   INCUBATOR COMPONENTS 

 

   

Control panel and keypad

 

 
 
 
 
 
 

 

 
 
 
 
 
 
 

 
 

1. Digital temperature indicator

 (TEMPERATURE 

o

C):  Normally, this indicator shows the chamber 

temperature.  In the setting mode, it shows the set value of the chamber temperature.  If the self 
diagnostic function detects any abnormality, an error code will be displayed. 

2. Heater lamp

 (HEAT):    This lamp lights when the heater is energized. 

3. Digital CO

2

 density indicator

 (CO

2

    %):    Normally, this indicator shows the CO

2

 concentration in the 

chamber.    In the setting mode, it indicates the set value of the CO

2

 concentration. 

4. CO

2

 inject lamp

 (INJECT):    This lamp lights when CO

2

 gas is being injected. 

5. UV indicator 

(UV):  This lamp lights when the UV lamp is ON [when an optional component 

MCO-80UVS (UV system set) is installed]. The blink of this indicator recommends the replacement of UV 
lamp. 

6. High humidity mode indicator

 (RH HIGH):  This lamp lights when the high humidity mode is 

activated. See page 21 for changing to the high humidity mode. 

7. Door lamp

 (DOOR):    This lamp lights when the outer door is open.

 

8. Water level alarm lamp 

(RH PAN):  This lamp blinks when the water in the water reservoir is less 

than approximately 5 L. 

9. Over heat lamp

 (OVER HEAT):    This lamp lights when the chamber temperature reaches the upper 

limit set value. It starts to blink when the chamber temperature is back in below the upper limit set value. 

10. Upper limit regulator

:    This regulator is used to set the upper temperature limit. 

11. Set key

 (SET):    Pressing this key to enter the setting mode, and the digits to be set will blink. 

12. Calibration key

 (CAL):    By pressing this key for approximately 5 seconds, the unit enters calibration 

function mode. 

13. Alarm buzzer stop key

 (BUZZER):    Press this key to silence the buzzer when the alarm operates 

and the buzzer sounds.

 

14. CO

2

 gas supply line indicator 

(A/B):  The lamp for the supply line currently in use lights up 

provided that MCO-80GC gas auto changer is installed. 

15. CO

2

 gas supply line switching key

 :    This key to select CO

2

 gas supply line is available only when 

a gas auto changer MCO-80GC (option) is installed.  When one CO

cylinder is empty, the CO

is 

supplied by the other cylinder automatically.

 

2 3 

10 

11 

12 

13 

14 

15 

17 

18 

19 

16 

Summary of Contents for MCO-80IC Series

Page 1: ...80IC Series Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 42 for model number Operating Instructions CO2 Incuba...

Page 2: ......

Page 3: ...CAUTIONS FOR CULTURE 18 START UP OF UNIT 19 SETTING OF CHAMBER TEMPERATURE AND CO2 DENSITY 20 KEY LOCK FUNCTION 23 ALARMS SAFETY FUNCTIONS 24 SETTING OF ALARM RESUME TIME 26 Operation after power fai...

Page 4: ...ating instruction are subject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instruction is lost or the pag...

Page 5: ...Symbol shows this symbol means caution this symbol means an action is prohibited this symbol means an instruction must be followed Be sure to keep this operating instruction in a place accessible to u...

Page 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 7: ...he power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply p...

Page 8: ...e to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet copy the last page when you request any repair or maintenance fo...

Page 9: ...idity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations not to exceed 10 of the nominal voltage Other supply voltage fluctuations...

Page 10: ...8 INCUBATOR COMPONENTS Rear upper side Upper left side Bottom of chamber When the bottom cover is removed 1 2 3 4 5 6 inner face of ceiling 7 8 9 11 10 12 13 14 15 16 17 18 19 20 22 23 21 25 24...

Page 11: ...B are available If MCO 80GC is not used only A is available Refer to page 15 for gas cylinder connection Ensure that the gas pressure is set at 0 1 MPa G 1 0 kgf cm2 G 14 5 psi G Refer to page 39 for...

Page 12: ...lights when the outer door is open 8 Water level alarm lamp RH PAN This lamp blinks when the water in the water reservoir is less than approximately 5 L 9 Over heat lamp OVER HEAT This lamp lights whe...

Page 13: ...in the controller 18 Numerical value shift key Pressing this key in the setting mode causes the numerical value to shift In key lock mode pressing this key makes key lock ON or OFF 19 Digit shift key...

Page 14: ...n between COM to N C Close between COM to N O Contact capacity DC 30 V 2 A Note When the power switch is OFF or the power failure condition the contact output between COM to N O is CLOSE the contact o...

Page 15: ...ll the unit in stable condition to avoid the vibration or noise Unstable condition may cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the inst...

Page 16: ...unterclockwise to contact them to the floor Ensure the unit is level 3 Fix the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall with these hooks and rope or chain 4 Gr...

Page 17: ...nect the gas regulator to the connecting port for CO2 gas pipe located at the rear upper side of the CO2 incubator 3 Set the CO2 pressure on the secondary side to 0 1 MPa G 1 0 kgf cm2 G 14 5 psi G at...

Page 18: ...ut abnormal temperature rising If there is too much heat emission from the equipment in the chamber The chamber temperature may deviate very much from the setting temperature when a heat generator hea...

Page 19: ...ially when the culture medium is spilled clean and disinfect the chamber immediately Refer to page 27 Routine maintenance for details Keep the inside panels dry To protect the inside of the unit from...

Page 20: ...ture distribution and CO2 concentration in the chamber Stored materials Never place acid or alkaline materials or materials that release corrosive gas in the chamber Such materials can cause failure r...

Page 21: ...the chamber and internal attachments Refer to Sterilizing of chamber and attachment on page 27 3 Place the drain tank under the unit Insert a drain outlet into the hole of drain tank 4 Fill the water...

Page 22: ...digit shift key and numerical value shift key set the figure to 05 0 When pressed the changeable digit is shifted When pressed the figure of settable digit changes 6 Press enter key ENT Set CO2 densi...

Page 23: ...s enter key ENT Set value is memorized The current chamber temperature is displayed Table 2 No humidifying mode setting procedure Description of operation Key operated Indication after operation 1 The...

Page 24: ...he unit is under normal conditions The unit enters calibration mode by pressing this key Wrong key operation may affect the basic performance When the unit enters calibration mode by mistake wait for...

Page 25: ...g change from key lock OFF to key lock ON Description of operation Key operated Indication after operation 1 The current chamber temperature and CO2 density are displayed 2 Press digit shift key for 5...

Page 26: ...ternately with the temperature on the digital temperature indicator Gas supply line indicator blinks Intermittent tone Gas supply line is altered Remote alarm Chamber temperature sensor abnormality If...

Page 27: ...is displayed alternately with the temperature on the digital temperature indicator Intermittent tone Remote alarm Glass frame heater abnormality If the glass frame heater or the glass frame heater rel...

Page 28: ...tal CO2 density indicator The middle digit blinks 5 Set the figure to 020 with the numerical value shift key When pressed the figure of settable digit changes 6 Press enter key ENT The setting is memo...

Page 29: ...ff lightly Sterilizing of chamber and attachments When the chamber of the unit is contaminated the chamber and internal attachments should be cleaned and sterilized as follows Note Take care not to da...

Page 30: ...water reservoir with a gauze containing alcohol for sterilization and then wipe off with a dry gauze 10 Wipe the water level sensor with a gauze containing alcohol for sterilization and then wipe off...

Page 31: ...blinking add approximately 15 L of preheated sterile distilled water at 37 o C See Fig 3 It might take approximately 2 minutes until the lamp goes off 4 Make sure the water level alarm lamp RH PAN on...

Page 32: ...gital CO2 density indicator goes out 3 Adjust the set value to the actual value of 36 8o C with the digit shift key and numerical value shift key then press enter key ENT 4 The unit automatically reve...

Page 33: ...en press the enter key ENT 6 The unit automatically reverts to the display mode Note This calibration is available when the setting of CO2 density is 2 or more Example For an internal CO2 density of 5...

Page 34: ...as in the cylinder Replace the cylinder soon While an equipment are operated in chamber Generation of heat by the equipment raises temperature in chamber If the chamber temperature is not equal to the...

Page 35: ...33 Note If the malfunction is not eliminated after checking the chart in the preceding page or the malfunction is not shown in the chart contact our sales representative or agent...

Page 36: ...d in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely The disposal of the unit should be accomplished by appropriate personnel...

Page 37: ...nformationen zur Entsorgung von elektrischen und elektronischen Ger ten private Haushalte Dieses Symbol auf Produkten und oder begleitenden Dokumenten bedeutet dass verbrauchte elektrische und elektro...

Page 38: ...ter les autorit s locales ou votre revendeur afin de conna tre la proc dure d limination suivre Spanish Informaci n sobre la eliminaci n para los usuarios de equipos el ctricos y electr nicos usados p...

Page 39: ...orma es sobre o m todo de elimina o correcto Italian Informazioni per gli utenti sullo smaltimento di apparecchiature elettriche ed elettroniche obsolete per i nuclei familiari privati Questo simbolo...

Page 40: ...act op met de lokale overheid of uw leverancier en vraag wat de juiste verwijderingsmethode is Swedish Information om kassering f r anv ndare av elektrisk elektronisk utrustning privata konsumenter Om...

Page 41: ...CO2 gas supply line A or B 6 When one cylinder is empty the indicator blinks buzzer sounds and E01 and current chamber temperature are displayed alternately on the digital temperature indicator while...

Page 42: ...ir decreases less than approximately 5 L water is supplied little by little to prevent the water temperature from dropping rapidly The water supply valve is opened and closed repeatedly Automatic wate...

Page 43: ...igital display CO2 controller PID control system CO2 density display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or more Water level sensor Thermal type Alarm High L...

Page 44: ...variation 0 15 ambient temperature 25 o C setting 37 o C 5 no load Chamber humidity Normal mode Over 80 R H High humidity mode Over 90 R H Maximum heat emission 6159 6 kJ h Usable environment conditio...

Page 45: ...tion of the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below...

Page 46: ......

Page 47: ......

Page 48: ...Printed in Japan LDCL041700 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...