CAUTIONS FOR CULTURE
●
Do not subject to direct air flow
Do not allow the air for air conditioning to hit the unit or door directly. Direct hit may cause condensation
or contamination.
●
Tray
Make sure to set the bottom tray to the bottom position of side duct with tray support. This may cause
condensation on bottom cover by varying air flow inside chamber.
●
Do not block holes on the side duct with the culture vessels.
When storing cultures in the chamber, take care not to block holes on the side duct with the Petri dishes
or roller-bottle racks to allow adequate air circulation. Blockage of holes may result in uneven
temperature distribution and CO
2
concentration in the chamber.
●
Stored materials
Never place acid or alkaline materials or materials that release corrosive gas in the chamber. Such
materials can cause failure resulting from discoloration or corrosion.
●
Open/close the doors gently
Ensure you close the doors gently. Robust closing may cause spillage of medium, incomplete closing,
or damage of gasket.
●
Fix the tray supports and trays securely
Incomplete installation may cause injury or damage.
●
Never lean or press on the glass.
Intentional force may cause injury if the glass breaks.
●
Do not lean on the door
This may cause injury, current leakage, or electric shock if the unit tips over or door drops out.
●
Alarm
Always investigate the cause and fix the alarm condition immediately when the alarm is activated. Refer
to page 24 for alarm details.