64
DRY HEAT STERILISATION
12.
On the
Sterilisation Step1
screen, the system
check starts automatically. If the system is normal,
the display leads the
Sterilisation Step2
screen.
Note:
If the system is abnormal, the display leads
the
Sterilisation Step6
screen. Resolve the cause of
the error displayed in the message display field, in
reference to Table 8 on page 89~90.
Press the
OK
key to return to the
Top
screen, and
perform dry heat sterilisation again.
13.
On the
Sterilisation Step2
screen, by holding
the
Ready to Start
slide key and sliding it right,
change to Yes. Press the
OK
key to lead the
Sterilisation Step3
screen and dry heat sterilisation
is started. Dry heat sterilisation is performed
automatically from Step3 to Step6 (procedure
14
).
·
Press the
Cancel
key to return to the
Top
screen.
Note:
·
The outer door is locked with electric lock for safety until completion of dry heat sterilisation.
·
It takes approximately 11 hours until completion of dry heat sterilisation.
WARNING
Do not use the unlock key to unlock the outer door during dry heat sterilisation even if a power
failure occurs.
Doing so may cause burns.
CAUTION
The electric lock will remain locked if a power failure occurs during dry heat sterilisation. After recovery
from the power failure, the cooling process (Step5) will start execution and finish automatically. Perform dry
heat sterilisation again because dry heat sterilisation is not completed.