11
1.
Outer door:
The outer door is held to the frame with the magnetic seal. The door heater is installed in
the door panel. The door opening is reversible. Contact our sales representative or agent to change the
door hinge from left to right or vice versa.
2. Inner door:
The inner door is made of tempered glass. However do not subject the glass to excessive
impacts.
3. Tray catches:
Insert tray to fit the concave portion on chamber.
4. Fan cover:
The fan cover serves as the inlet for circulating air. It is removable.
5. Duct:
The duct for the path for circulating air. It is removable.
6. Fan (inside the duct):
It can be sterilised in an autoclave.
7. UV lamp:
This UV lamp does not generate ozone. Never look directly at the UV light. Refer to page
74~78 for using. For replacement, contact our sales representative or agent.
8. Humidifying pan cover:
This cover prevents the UV light entering the chamber. Always use it. Using
without it may have a bad influence on the chamber temperature distribution and humidity recovery.
9. Humidifying pan:
Fill the humidifying pan with sterile distilled water, and set the humidifying pan with
the inner side flush against the back. Install the humidifying pan in a longitudinal direction as its shorter
side is placed in the back.
10. Door switch:
Detects the door opening/closing and stops the fan and electromagnetic valve for CO
2
when the door is open. The UV lamp* is also deactivated by the door opening.
11.
Key hole:
This is the hole to unlock with unlock key while outer door is locked by electric lock.
12. Switch cover:
Prevent the accident of gas tube disconnected or power off by the unexpected touch.
13. Power supply cord cover plate:
This plate is to prevent the removable power supply cord being
come off.
14. Connecting port A for CO
2
gas pipe:
Refer to page 20 for gas cylinder connection. Ensure that the
secondary side pressure of the gas regulator is the specified value (refer to page 20 or 79).
Note:
When the optional MCO-21GC gas auto changer is installed, both ports A and B are available. Refer
to page 79~82 for gas auto changer.
15. Glow starter:
The glow is started for the UV lamp.
16. Power switch:
This is the main switch for the incubator (ON-“I”, OFF-“O”). It also functions as an
overcurrent breaker.
17. Sample air outlet:
The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
18. Service port:
It is possible to relocate the sample air outlet from
17
to here. Contact our sales
representative or agent.
19. USB port:
Insert USB memory to export operation, alarm and dry heat sterilisation logs. Refer to page
45~57.
Note:
It is impossible to use USB memory which is required password input.
20. Access port:
Place the silicon caps on both outside and inside of the port when the port is not being
used.
21. Remote alarm terminals:
This terminal informs the alarm to remote location by connecting to external
alarm unit. Refer to page 14.