56
UV LAMP (OPTION)
Setting UV lamp ON period
To change the UV lamp lighting times, follow the procedure below.
Lighting time setting range: 0 - 30 minutes
Default setting (factory setting): 10 minutes
Note:
・
The recommended UV lighting time (UV TIMER) is 10 minutes. Less than 10 minutes may have an
insufficient bactericidal effect.
・
If it is set to 0 minutes, the UV lamp will not light up.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. Move the cursor on the Menu screen to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the Alarms & Controls screen.
3. Move the cursor on the Alarms and Controls screen to UV Setting [Fig.
2] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the UV Setting screen.
4. With the cursor on the UV Setting screen on Lighting Time [Fig. 3],
press the enter key (ENTER).
▶
The right of the display will change to the UV Lighting Time screen,
and the current setting value (10) will be displayed [Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the UV
Setting screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key
on the control panel to return to the UV Lighting Time screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set ▼
UV Lighting Time
Invalid Value
Press Any Key
0min – 30min
<Right side of the display>
<Right side of the display> [Fig. 4]
UV Lighting Time
10
min
0min – 30min
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display> [Fig. 3]
UV Setting
UV Lighting Time
UV Life Counter
Auto-Extended Time
UV 24h Mode Start