52
ALARM PARAMETERS
Setting door alarm delay
When the incubator is in the alarm state of door, the alarm buzzer will sound after the set time of door
alarm delay passed.
Settable range: 1 minute~30 minutes
Factory setting: 2 minutes.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. Move the cursor on the Menu screen to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the Alarms & Controls screen.
3. Move the cursor on the Alarms & Controls screen to Other Alarm Set
[Fig. 2] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the Other Alarm Set screen.
4. When the cursor on the Other Alarm Set screen is on Door Alarm
Delay [Fig. 3], press the enter key (ENTER).
▶
The right side of the display will change to the Door Alarm Delay
screen, and the current setting value (02) will be displayed
[Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the
Other Alarm Set screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key
on the control panel to change to the
Door Alarm Delay
screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Note:
When the incubator is recovered from the alarm state within the set time of door alarm delay, the
buzzer doesn’t sound after the elapse of the set time of door alarm delay.
<Left side of the display> [Fig. 3]
Other Alarm Set
Door Alarm Delay
Ring Back Delay
Remote Alarm
Door Alarm Delay
02
min
1min – 30min
<Right side of the display> [Fig. 4]
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
Door Alarm Delay
Invalid Value
Press Any Key
1min – 30min
<Right side of the display>
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set ▼