34
BASIC PARAMETERS
Setting CO
2
density
Set the CO
2
density normal operation according to the following procedure. The incubator automatically
starts operation using these settings after power-on.
Settable range : 0.0
%〜
20.0
%
Factory setting : 0.0
%
Note:
·
When operating the incubator for the first time or after not using it for an extended period of time, operate
it for at least about 8 hours until the chamber temperature and the CO
2
sensor are stable after setting the
chamber temperature to the desired temperature and setting the CO
2
density to 0 %. Then change the
setting to the desired CO
2
density.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display will change to the Set Value screen.
3. Move the cursor on the Set Value screen, to CO
2
Setting [Fig. 2] using
the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the display will change to the CO
2
Setting screen,
and the current setting value (05.0) will be displayed [Fig. 3].
4. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
5. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the Set
Value screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key
on the control panel to change to the CO
2
Setting screen [Fig. 3].
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Right side of the display> [Fig. 3]
CO
2
Setting
05.0
%
0.0
%
-
20.0
%
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display> [Fig. 1]
Set Value
Temperature
: 37.0°C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
Overtemp
: 53.0°C
<Left side of the display> [Fig. 2]
CO
2
Setting
Invalid Value
Press Any Key
0.0% – 20.0%
<Right side of the display>