Friabilimeter
operating instructions
Page 42 of 48
Revision 7
© 2016 – Pfeuffer GmbH
NOTE
If the chain and chain sprockets are heavily contaminated, rub them clean with a
dry, clean and lint-free cloth.
Grease the components lightly with a spindle and anti-friction bearing grease.
Place the chain back on the chain sprocket.
Position the new motor and the mounting plate on the motor pin using the three M4 fillister head
screws (with lock washers and plain washers).
NOTE
Do not overtighten the chain.
Otherwise, the motor and bearings might be damaged.
Allow the chain to sag slightly.
Screw the motor on firmly.
Insert the motor plug, see
Figure 33.
Attach the front plate using the four outer countersinking screws. Pay attention to the correct
installation position of the two upper serrated washers, see
Figure 31.
Screw the pull handle in firmly.
Insert the sieve drum and put on the front hood.
After renewal of the drive rolls, check the device setting with test malt. It may be necessary to adjust
the settings, see
chapters 8.5
and
Make a note of the renewal, the calibration and the subsequent check with test malt in a test log
(see master copy on last page).
Checks
9.6
At the end of the work, check the following:
The work carried out is complete.
Check the wiring in the housing for kinks, chafing or charred points.
Damage on the covers or insulation.
Check that no tools have been left in or on the machine.
All subassemblies function correctly in setup or manual mode.
If all functions are correct, the machine can be handed over to the owner.
NOTE
Following cleaning, maintenance or exchanging wearing parts, check that all safety
devices are functioning correctly.