Friabilimeter
operating instructions
© 2016 – Pfeuffer GmbH
Revision 7
Page 29 of 48
Precise adjustment using the setscrew
8.5
Before you make a precise adjustment, carry out and make a note of an investigation
with a test malt (see master copy on last page).
Disconnect the mains cable from the electrical power supply, or pull out the
coupler.
There is a plastic cap on the top of the FRIABILIMETER. Under this plastic cap, there is a setscrew inside
the housing.
This setscrew directly acts on the spring travel and thus on the force exerted on the sieve drum by the
pressure roll.
Figure 26: Setscrew
Item Designation
1
Plastic cap
setscrew
2
Pull handle
Remove the plastic cap.
Engage the pull handle in the top position.
Adjust the setscrew using a flat-blade screwdriver.
Turn the setscrew
clockwise (+) to increase the result.
Turn the setscrew
counterclockwise (-) to reduce the result.
Make a note of the precise setting and the subsequent check with test malt in a test log (see master
copy on last page).
One rotation corresponds to about 0.1 to 0.2 %.
At a lower friability, the effect on the investigation result is generally greater than at
a higher friability.