Operating Instructions
MILOMAT
laboratory mill
© 2017 Pfeuffer GmbH
Revision 1
Page 27 of 32
Figure 27: Remove the eddy current disc
Remove the eddy current disc by means of the two
lens head screws M3x8 and fan discs.
Figure 28: Remove/install the V-belt
Remove the old V-belt.
Install the new V-belt.
Fix the eddy current disc by means of the two lens
head screws M3x8 and fan discs.
Figure 29: Insert the motor
Insert the motor into the housing by placing it
diagonally on the V-belt.
Turn in the 4 motor fixing screws loosely (hexagon
socket screws M5x16, washers and spring washers).
Figure 30: Tension the V-belt
Tighten the V-belt by moving the motor plate
downwards. The correct belt tension is reached
when the belt can be pushed through approx. 10
mm with low effort.
Tighten the motor fixing screws properly (hexagon
socket screws M5x16, washers and spring washers).
Fasten the housing, see
chapter 8.4.