24
The Pfeiffer Vacuum parameter set
7.3.4
Control via interface
Configuration via parameters
[P:060]
and
[P:061]
.
7.3.5
Digital inputs on "X3"
Configuration via parameters
[P:062]
and
[P:063]
.
7.4
Operation with the Pfeiffer Vacuum parameter set
7.4.1
Factory settings
The electronic drive unit is pre-programmed in the factory. This guarantees proper, reli-
able turbopump operation without the need for additional configuration.
7.4.2
Checking the adjustments
Before operating with parameters, check set values and control commands for their
suitability for the pumping process.
7.4.3
Gas type dependent operations
Friction causes the rotor to heat up severely under gas load and high rotation speed. To
avoid overheating, the electronic drive unit has implemented power-rotation speed-char-
acteristics, whereby the pump can be operated at every rotation speed with the maxi-
mum allowable gas load without danger of damage. The maximum power consumption
depends on the gas type. Three characteristics are available in order to completely ex-
haust the pump's capacity for each gas type.
6 = Pressure value 1
Pressure value signal;
0 V: error
1 V: underrange
1,5 - 8,5 V for sensor RPT p (hPa) = 10
(U-5.5 V)
1,5 - 8,5 V for sensor IKT p (hPa) = 10
(U-10.5 V)
9 V: overrange
7 = Pressure value 2
8 = Control fore-vacuum
Fore-vacuum signal; control of Pfeiffer Vacuum
turbo pumping stations
Option
Description
Option [P:060]
Description
1 = remote
Operation via connection "
remote
"
2 = RS-485
Operation via connection "
RS-485
"
4 = PV.can
For service purposes only
8 = Field bus
Operation via field bus
16 = E74
Operation via connection "
E74
"
Option [P:061]
Description
0 = off
Interface selection via [P:060]
1 = on
Interface selection locked
Option
Description
0 = Deactivated
Connection deactivated
1 = Venting release
Control is equal to parameter [P:012]
2 = Heating
Control is equal to parameter [P:001]
3 = Sealing gas
Control is equal to parameter [P:050]
4 = Run-up time control
Control is equal to parameter [P:004]
5 = Rotation speed setting mode
Control is equal to parameter [P:026]
6 = Motor
Control is equal to parameter [P:023]
7 = HV sensor release
Control is equal to parameter [P:041] (only 0 or 1)