2.1 Safety precautions
• Do not loosen any plug connection during operations.
• Wait for the rotor to reach standstill before peforming work on the high vacuum
flange.
• When using sealing gas, the pressure in the hose connection must be limited to
2 bar via an overpressure valve.
• Keep leads and cables well away from hot surfaces (> 70 °C).
• Always ensure a safe connection to the protective earthing conductor (PE, pro-
tection class I).
• Discrete operating of the turbopump and the electronic drive unit is only allo-
wed after authorisation by Pfeiffer Vacuum. In this case the turbopump must be
connected to the PE.
• The danger of an electrical shock in the case of ground leakage must be elimi-
nated immediately (red LED flashes).
• Never fill or operate turbopump with cleaning agent.
2.2 Proper use
• Only operate the turbopump with a prooved backing pump.
• Only operate the turbopump with TC by a specified Pfeiffer Vacuum power sup-
ply. The use of other power units than the intended, is only permitted after con-
sultation with Pfeiffer Vacuum.
• Only operate the turbopump with TC 600 in ambient conditions up to protection
class IP 30.
• Only operate the turbopump together with a water cooling unit or in ambient
conditions up to protection class IP 54 after installing a specified TC cover plate.
NOTE
Duty to inform
Each person involved in the installation, operation or maintenance of the vacuum
pump must read and observe the safety-related parts of these operating instructions.
Î
Absolute observe the safety information for vacuum pumps (PT 0300 BN) !
Î
The operator is obligated to make operating personnel aware of dangers originating
from the vacuum pump, the pumped medium and the entire system.
NOTE
Installation and operation of accessories
Pfeiffer Vacuum pumps can be equipped with a series of adapted accessories. The
installation, operation and maintenance of connected devices are described in detail in
the operating instructions of the individual components.
Î
For information on the operating instructions of components, see "Accessories".
Î
Use original accessory parts only.
NOTE
CE conformity
The manufacturer's declaration becomes invalid if the operator modifies the original
product or installs additional components!
Î
Following installation into a plant and before commissioning, the operator must
check the entire system for compliance with the valid EU directives and reassess it
accordingly.