36
Technical data and dimensions
Run-up time
1.6 min
1.6 min
1.6 min
Ultimate pressure with OnTool
TM
DryPump
< 1 · 10
-7
hPa
< 5 · 10
-10
hPa
< 1 · 10
-7
hPa
Rotation speed ± 2 %
49200 rpm
49200 rpm
49200 rpm
Rotation speed: variable
65-100 %
65-100 %
65-100 %
Power characteristic line in gas mode 1, vertex A
150/49200 W/rpm
150/49200 W/rpm
150/49200 W/rpm
Power characteristic line in gas mode 1, vertex B
175/42000 W/rpm
175/42000 W/rpm
175/42000 W/rpm
Power characteristic line in gas mode 0, vertex C
135/49200 W/rpm
135/49200 W/rpm
135/49200 W/rpm
Power characteristic line in gas mode 0, vertex D
140/42000 W/rpm
140/42000 W/rpm
140/42000 W/rpm
Power characteristic line in gas mode 2, vertex E
245/49200 W/rpm
245/49200 W/rpm
245/49200 W/rpm
Power characteristic line in gas mode 2, vertex F
320/42000 W/rpm
320/42000 W/rpm
320/42000 W/rpm
Sound pressure level
≤
50 dB (A)
≤
50 dB (A)
≤
50 dB (A)
Relative humidity of air
5-85, non-condensing %
5-85, non-condensing %
5-85, non-condensing %
Protection category
IP 54
IP 54
IP 54
Connection pressure max. for venting/sealing gas valve 150 kPa
150 kPa
150 kPa
Operating voltage
48 (± 5 %) V DC
48 (± 5 %) V DC
48 (± 5 %) V DC
Operating voltage power supply
90-265 V AC
90-265 V AC
90-265 V AC
Integral leak rate
< 1 · 10
-7
hPa l/s
< 1 · 10
-7
hPa l/s
< 1 · 10
-7
hPa l/s
Power consumption max.
420 W
420 W
420 W
Current consumption max.
8.75 A
8.75 A
8.75 A
Shipping and storage temperature
-25-+55 °C
-25-+55 °C
-25-+55 °C
Mounting orientation
in any orientation
in any orientation
in any orientation
Venting connection
G 1/8"
G 1/8"
G 1/8"
Weight
12.7 kg
18.9 kg
13.5 kg
Cooling method, standard
Water
Water
Water
Cooling method, optional
Air
Air
Air
Cooling water temperature
15-35 °C
15-35 °C
15-35 °C
Cooling water consumption
100 l/h
100 l/h
100 l/h
Permissible magnetic field max.
6 mT
6 mT
6 mT
Interfaces
RS-485, Remote
RS-485, Remote
RS-485, Remote
Parameter
HiPace
TM
800 P
HiPace
TM
800 P
HiPace
TM
800 P