Order Parameter
Description
Functions
Min Max
SBY
none
Switches from stand-
by speed to the de-
fault setting
Stand-by speed is reset at 35 Hz.
ACP
ON or OFF Start/Stop the pump
The pump rotation starts with ACPON, pump stops with
ACPOFF.
*Only rotation speeds recommended in the settings table are allowed
(see chapter “Setting of the rotation
Order Pa-
rame-
ter
Description
Functions
STA
none
Pump status
Example: #adr,xxxxxx yyyyyy,zzzzzz,sssss,iiii,www,ppp,vvv,ttttt<CR>
● sssss = pump rotation speed (min
-1
)
● iiii = motor power (W)
● www = reserved (by default 000 )
● ppp = frequency converter tem-
perature (°C)
● vvv = reserved (by default 000)
● tttt = pump operating time (from
first start-up) (h)
xxxxxx: sta-
tus bits
Bit 5
4
3
2
1
0
0
0
0
0
0
set speed set
point
0
0
1
-
-
1
pump run-
ning
1
nominal
speed set
1
stand-by
speed reached
-
yyyyyy: fault
bits
Bit 5
4
3
2
1
0
0
-
0
0
0
0
0
1
1
power tran-
sistor not
powered
(red LED
copy = 1)
1
motor tem-
perature
too high
1
motor cur-
rent too
high
-
-
-
zzzzzz: alert
bits
Bit 5
4
3
2
1
0
0
0
0
0
0
0
OFF
1
-
-
-
-
-
-
Installation
28/56