Example:
Header character
Pump address
Order
Parameter
End character
#
ADR
ODR
XXXX
<CR>
The responses
Header character
Pump address
Order
End character
#
ADR
yyyxxxabc
<CR>
Interpretation of the responses
OK
If everything is OK, or specific response to the order sent
ERR0
Setting fault
ERR1
Context fault
ERR2
Parameter fault
ERR3
Order fault
Example of dialog
Order
#005ACPON<CR>
Response
#005,OK<CR>
5.6.3 Command list
Order Parameter
Description
Functions
Min Max
ADR
xxx
Number given to the
pump in the serial link
Can be used only when the pump is stopped.
● adr = pump address before change
● aaa = new pump address
When the address of the product is unknown, it is possible
to recover the pump address only if this pump is connected
in the link.
For this, send the order:
#???ADR<CR>
??? chain of three ASCII characters.
000
255
IDN
none
Identification of the
connected product
Example: #adr,ACP 28- VB.05<CR>
Returns the type of pump connected to the computer
(12345), the software version (V) followed by the software
release (zz).
NSP
none
Switches from set
point speed to the
nominal speed
Example: #adrNSP<CR>
● Nominal speed =
80 Hz
RPM
nnnn*
Set point speed set-
ting
Example:
#adrRPMnnnn<CR>
Set point speed can be set in steps of 10 min
-1
:
● Minimum speed:
35 Hz
for ACP 28 - 40
● Maximum speed:
80 Hz
for ACP 28 - 40
Note:
Before changing the set point speed with the RPM or-
der, the SBY order must be sent.
35
80 Hz
Installation
27/56