5.5.5 RESET (
PRESS AND HOLD RESET KEY WHILE PRESSING ON RECORD
)
This option re-establishes the default settings of the instrument.
The default settings of the instrument consist of:
ANALYZER CONFIG:
Frequency: 60Hz
Full scale of the clamps: 1000A
Type of electrical system: 4 wires
RECORDER CONFIG:
Start: Manual (the recording is started at 00 sec mark on clock after pressing the RECORD key)
Stop: Manual
Integration period: 5S
Recording of harmonics: ON
Recording of Sag and Surge: OFF
Voltage Reference for Sag and Surge detection: 480V
Upper Limit for Sag and Surge detection: 10%
Lower Limit for Sag and Surge detection: 10%
Selected voltages: V1, V2, V3
Selected voltage harmonics:
Selected currents: I1, I2, I3, IN
Selected current harmonics:
Powers, Pf and cos
Θ
Θ
selected:
Pt, P1, P2, P3
Qt, Q1, Q2, Q3
Qt, Q1, Q2, Q3
St, S1, S2, S3
Pft, Pf1, Pf2, Pf3
dpft, dpf1, dpf2, dpf3
Energies:
Eat, Ea1, Ea2, Ea3
Erit, Eri1, Eri2, Eri3
Erct, Erc1, Erc2, Erc3
Est, Es1, Es2, Es3
Note: The RESET command will not erase the instrument’s memory.
29
DM-4 Power Quality Recorder