T3 / T3M Instruction Manual
127
charging.
5.
Scanning the energy on the Kinetics
mode, replace with a new tungsten
lamp if the energy fluctuates.
4.
The menu flashes after
turning on the
instrument.
1.
Check if the power supply is stable or
the power for the battery is enough.
2.
Turn on the instrument again and run
the software.
5.
Baseline flatness is over
the limitation.
1.
Do the baseline correction again.
2.
Ensure that the cell holder is fixed.
3.
Check if the power for the battery is
enough.
4.
Check if the energy is too low, to do
the energy balance.
6.
The drift of the
instrument is big.
1.
Check if the power for the battery is
enough.
2.
Repeat zeroing after warming up.
7.
The wavelength of the
instrument is not precise.
Return the instrument to the
manufacturer for servicing.
8.
It suddenly shut down
1.
Check if the power for the battery is