T3 / T3M Instruction Manual
126
Chapter 7 Troubleshooting
No.
Phenomenon
Trouble-shooting
1.
The touch-screen is not
accurate
1. Restart
2. Re-orientate the touch-screen
2.
The energy from the light
source is low.
1.
Energy balance.
2.
Tungsten lamp and power switch
broken.
3.
Very low energy and bad baseline,
replace with a new tungsten lamp.
4.
Check if there is something block the
light in the cell holder.
3.
The noise of the
instrument is too large.
1.
Energy balance, blank correction.
2.
Check if the sample is turbidity and if
there is besmirch in the sample
chamber for the fiber.
3.
Check if the power for the battery is
enough.
4.
Check if the power supply is stable
when using charger for battery