Error handling
This section contains information for handling errors given by the TriNEST Microplate incubator and shaker instrument.
General
Due to the very compact design of the 1296-0050 TriNEST Incubator Shaker it is very limited what is cost efficient to
repair on the unit. Therefore the policy is for most failures to fully replace the instrument with a new one.
Quality check of shaking and heating systems are built in the instrument and in case the instrument displays one of the
below error messages, reboot the instrument and if the problem persists, contact authorized service person.
Error 1, Tmp sens
n
Temperature sensor number
n
cannot be found.
Error 2, Heater
n
Heater element number
n
is detached or has short circuited.
Error 3, Motor belt
Oven is not moving, conveyer belt has been detached, speed indicator sensor is detached or broken, or there is an
error with motor.
Error 4, Motor rpm
The shaking rpm of the oven is incorrect. This can be caused by error in conveyer belt or in motor.
TriNEST Microplate incubator and shaker
Error handling
24