Shaking time: 0 - 48 h in increments of 1 min.
Interval shaking active periods: 3 s - 60 min. in increments of 1 s or 1 min. If set to 0 s, shaking is off.
Note:
Although the instrument will accept a value shorter than 3 seconds for a shaking active period, it may cause
errors and therefore must not be used.
Note:
Interval shaking rest period: 0 s - 60 min. in increments of 1 s or 1 min.
All three incubation chambers are subject to the same shaking conditions.
Time
Incubation time: up to 48 hours in steps of 1 minute.
Parameter sets
Up to 20 parameters sets (programs) can be stored in the instrument.
Measurements
The measurements of the TriNEST Microplate incubator and shaker instrument are as follows.
Physical dimensions
Depth: 350 mm (13.8 inches)
Width: 388 mm (15.3 inches)
Height: 225 mm (8.9 inches)
Weight
One unit: 15.7 kg (34.6 lb)
Three stacked units: 47.1 kg (103.8 lb)
Environmental conditions
The environmental conditions of the TriNEST Microplate incubator and shaker instrument are as follows.
Operating temperature
Ambient temperature range during operation: 15 - 35°C
Humidity
Relative humidity during operation: 10 - 75%
Altitude
Altitude: 2000 m
TriNEST Microplate incubator and shaker
Specifications and features
20