Operating instructions
Switch on power with the switch at the back of the instrument. The software and hardware version numbers will briefly
appear followed by Program 01.
Program selection
The current parameter values for Program 01 will be
displayed. To see another program, turn the selection
wheel until the number of that program is displayed.
Parameter editing
After selecting the program number, press
the
Right arrow
for
three
seconds. The
first character of the program name will be
highlighted for editing.
Note:
During parameter editing, press
the
Right (or Left) arrow
to accept the
selection and move the cursor to the next
(or previous) parameter.
Program name
- turn the selection wheel
either clockwise or counter-clockwise until
the required character appears. Press the
Right arrow.
The cursor will move to the
next position.
Note:
The order of the characters that appear when you turn the selection wheel is:
• capital letters (A-Z)
• small letters (a-z)
• numbers (0-9)
• symbols (:!"#$%&()*+,-./_space) "Space" is the last symbol before "A".
When you press the
Right arrow
after the
sixth position, the cursor will move to the
incubation temperature value.
Set the following parameters using the
selection wheel and the arrow buttons:
TriNEST Microplate incubator and shaker
Operating instructions
14