84 . Spotlight 150 User's Guide
4.
Lay the assembly, crystal up, on a clean work surface. Remove the coned knurled collar
to expose the crystal holder.
5.
Carefully grab the crystal holder and remove it from the ATR assembly. Set the crystal
holder aside and be careful not to damage the crystal.
6.
Look down into the center of the crystal bearing, notice the compression spring and the
plunger retracted and off to one side. Spray methanol into the bore and allow it to drain
from the assembly.