78 . Spotlight 150 User's Guide
3.
Tighten the knurled thumb-nuts.
4.
Move the microscope slide to a fresh area of tape.
5.
Lower the crystal onto the tape.
6.
Retract the crystal.
7.
Look through the microscope at the impression of the crystal on the tape. Decide
whether the impression is in the center of the field of view.
8.
If necessary, repeat steps 1 to 7 until the ATR objective is aligned.
Checking the Infrared Alignment
The following procedure allows you to determine the energy reaching the detector when the
ATR crystal is in place. If the crystal is aligned correctly, the energy level measured should be
at least 20% of the maximum energy measured in reflectance mode; see
on page 75 for details.
1.
Place the slide holding the reference mirror (supplied with the microscope) on the
Sample Stage.
2.
Look at the video image on the Setup Microscope tab and focus the microscope on the
surface of the mirror.
There are usually dust particles or scratches on the surface that you can use to focus
on.
3.
Lower the crystal onto the reference mirror.
4.
Enter
100
in the
Width
and
Height
text boxes in the Apertures section of the Setup
Microscope tab.
5.
Click
on the Measurement bar and monitor the energy reaching the detector.
The energy level measured should be at least 20 of that measured in reflectance mode;
see
Measuring the Energy in Reflectance Mode
on page 75 for details.
6.
Click
.
Summary of Contents for Spotlight 150
Page 1: ...Spotlight 150 User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...Overview of the Spotlight 150 ...
Page 32: ...32 Spotlight 150 User s Guide ...
Page 33: ...Getting Ready to Use the Spotlight 150 ...
Page 45: ...Preparing Samples ...
Page 58: ...58 Spotlight 150 User s Guide ...
Page 59: ...Techniques for Collecting Spectra ...
Page 94: ...Maintenance ...
Page 102: ...Appendices ...