09931238D
30
a.
Start Temp (°C): Shows the starting temperature of experiment.
b.
Measurement Temp (°C): Shows the temperature at which the sample will be measured.
c.
SP Temp (°C): Shows the current temperature while it goes up/down to the measurement
temperature.
d.
Block Temp (°C): Show the set temperature to measurement temperature.
e.
Probe 1 Temp (°C): Shows the temperature of Probe 1.
f.
Probe 2 Temp (°C): Shows the temperature of Probe 2.
g.
Monitor: Monitored temperature is shown out of Block, Probe 1 and Probe 2.
h.
Hold time (min): Shows the set holding time.
i.
Elapsed time (m:s): Shows the elapsed time to reach the sampling temperature.
j.
Status: Shows the status of experimental progress.
k.
Error Range (°C): Shows the temperature tolerance between the sampling temperature and
monitored temperature. The measurement will be made only when the temperature tolerance
reaches within the set error range
NOTE
: To pull out the cell easily, use the cell lifting knob.
Figure 21 Using the cell lifting knob
11.
After the measurement is finished, the results are displayed in the result window. Save or print
results as required.