09931238D
25
a.
Start Temp (°C): Shows starting temperature of the experiment.
b.
Measurement Temp. (°C): Shows the temperature at which the sample will be measured.
c.
SP Temp (°C): Shows the current temperature while it goes up/down to the measurement
temperature.
d.
Block Temp (°C): The temperature of Cell Block.
e.
Probe 1 Temp (°C): The temperature of Probe 1.
f.
Probe 2 Temp (°C): The temperature of Probe 2.
g.
Monitor: Monitored temperature is shown out of Block, Probe 1 and Probe 2.
h.
Hold time (min): Shows the set holding time.
i.
Elapsed time (m:s): Shows the elapsed time to reach the sampling temperature.
j.
Status: Shows the status of experimental progress.
k.
Error Range (°C): Shows the temperature tolerance between the sampling temperature and
monitored temperature. The measurement will be made only when the temperature tolerance
reaches within the set error range.
NOTE:
To pull out the cell easily, use the cell lifting knob.
Figure 20 Pull out the cell
12.
After the measurement is finished, the results are displayed in the result window. Save or print
the results as required.
Thermal Denaturation
NOTE
: Start the sample measurement after warming up the system for at least 20 minutes.
1.
Launch the UV Lab software and then, select Thermal Denaturation in the Experiment Type
and click OK.